Rigel UNI-SiM
The new Rigel UNI-SiM is truly unique, incorporating 6 vital signs into one handheld simulator. Battery-operated, the UNI-SiM can simulate NiBP, SpO2, ECG wave forms, temperature, respiration and invasive blood pressure simultaneously
Further product information.
The UNI-SiM is the only simulator capable of undertaking six synchronised vital signs parameters tests, enabling medical device engineers to quickly, easily and accurately perform NiBP, SpO2, ECG, temperature, IBP and respiration functionality tests simultaneously using a single portable instrument.
The battery powered vital signs simulator reduces the time taken to test the correct performance of a wide range of medical devices and equipment used in hospitals, operating theatres and other facilities.
It synchronises the full functionality of Rigel's UNI-SiM and SP-SiM as well as a comprehensive patient simulator (ECG, invasive blood pressure, respiration and temperature) to cut simulation times and deliver cost-saving benefits - engineers no longer need to use a variety of different instruments for testing these functions separately.
Whilst the UNI-SiM incorporates a range of custom settings that include a variety of paediatric and adult NIBP pressures, pulse volume adjustments, heart rate and manufacturer-specific O-curves, the UNI-SiM is fully customisable to meet specific performance test conditions.
The UNI-SiM also utilises new and advanced technologies to maximise the accuracy of each simulation. For example, both electronic and optical SpO2 simulation methods are combined into a single test setup, and manufacturer specific simulation curves are included during NIBP simulation, reducing uncertainties thus improving the test time. Featuring Bluetooth connectivity, simulation test results can be stored within the instrument and printed wirelessly to the rugged battery operated Elite Test and Tag printer.
Key Features
■ 6 in 1 vital signs tester
Combining simulation of NIBP, SpO2, ECG, temperature, respiration and IBP
■ Synchronised heart rate
Test across all parameters from one centralised heart rate simulation
■ Hand-held
This portable and lightweight unit is ideal for field service technicians
■ Battery operated
Convenient and fast testing without mains power
■ Manual and automatic operation
Choose the operation mode, staying in full control of the test situation
■ Onboard data entry and storage
Store up to 10,000 asset records, making record management simple
■ Custom test protocols
Create your own automatic sequences to automate your PPM procedures.
■ Download / Upload software
Keep up to date with ease
■ Bluetooth communication
Transfer between PC and accessories
■ Barcode / Test and Tag capability
Onsite printing and scanning of barcode labels
UNI-SIM Specifications
1 Non-invasive Bloodpressure (NiBP)
- Simulation waveform Oscillometric
- Integrated Pump 0 to 350 mmHg user configurable
- Leak test user configurable between 0-350mmHg
- Chronometer configurable up to 999 secs
- Digital Manometer 0 - 410 mmHg
- Pressure Accuracy +/- 0.5% FS
- Pressure Units mmHg, inHg, kg/cm2, cmH2O, mBar, PSI,
- inH2O and kPa
2 Oxygen Saturation (SpO2)
- Range 50 to 100%
- Accuracy ± 1 bpm
- Chronometer test option to test response time SPO2 monitor
3 ECG Arrhythmia Simulator
- ECG full 12 lead simulation including high leveloutput
- Wave forms Normal Sinus Rhythm (NSR), ST Elevation, ST Depression, Myocardial Infarction, Tall T
- Rate 20 - 300 BPM typed value
- Performance waveforms Sine, Square, Triangle, and Pulse
- Pacer Waveforms Synchronous Atrial, Asynchronous Atrial
- Pacer Only Pacer Pulse rate 60 BPM, Ventricular Pacer rate 70 BPM, Atrial & Ventricular Pacer Heart rate 70 BPM, R Wave Detection
Arrhythmia Waveforms
- Atrial - Sinus Arrhythmia (SA), Missing beat, Atrial Flutter (AFLT), Atrial Fibrillation (AFB), Paroxysmal Atrial Tachycardia (PAT), Junctional Premature
- Contraction Atrial Conduction - First Degree AV Block , Second DegreeAV Block - Mobitz I, Second Degree AV Block Mobitz II, Third Degree AV Block , RightBundle Branch Block (RBB), Left Bundle Branch Block (LBB), Left Anterior Hemiblock
- Ventricular - Premature Ventricular Contraction - Intermittent Premature Ventricular Contraction - Continuous, Bigeminy, Trigeminy, Ventricular Flutter (VFLT)Ventricular Fibrillation (VFB), Ventricular Tachycardia (VTC), Right Focal PVC
4 Temperature
- Simulation YSI 400 / 700 Static
- Range preset at 25, 33, 37and 41°C
5 Respiration
- Rates 5, 10, 15, 30, 60, 120, 180 Breaths per Minute
- Base resistances 250, 500, 750, 1000 ohms
- Apnoea Simulation 0 - 60 seconds duration 0 - 300 seconds interval
6 Invasive Blood Pressure (IBP)
- Channels 2 channels
- Static 0 to 300mmHg. Typed values
- Dynamic 0 - 300mmHg for systolic & diastolic
7 General
- Operation Battery cell, insitu charge
- Battery charger 110-230VAC, 50/60 Hz
- Battery life 4-24 hours testing depending on simulations
- Memory Capacity Approx. 10,000 records
- Communication Via Blue Tooth
- Display Monochrome ¼ VGA full graphics
- Keypad Alpha-numeric
- Weight <1.5 kg
- Size (L x W x D) 270 x 110 x 75 mm / 10.5 x 4 x 3"
- Operating conditions 10-30'C, 0-90% RH - NC
- Storage environment -15' - +60'C
- Environmental Protection IP40